Although accurate volume measurements of solutions can be avoided in gravimetric methods of analysis, they are required for almost any other type of analysis involving solutions.

Volumetric Flasks

Volumetric flasks are used for the dilution of solutions to a certain volume. They come in a variety of sizes, from 1 mL to 2 L or more. A typical flask is shown in Figure 2.4. These flasks are designed to contain an accurate volume at the specified temperature (20 or 25°C) when the bottom of the meniscus (the concave curvature of the upper surface of water in a column caused by capillary action, see Figure 2.10) just touches the etched "fill" line across the neck of the glass. The coefficient of expansion of glass is small, and for ambient temperature fluctuations the volume can be considered constant. These flasks are marked with "TC" to indicate "to contain." Other, less accurate containers, such as graduated cylinders, are also marked "TC." Many of these are directly marked on the face by the manufacturer as to the uncertainty of the container measurement; for example, a 250 mL volumetric flask is "±0.24 mL," or roughly a 0.1% error.

Initially, a small amount of diluent (usually distilled water) is added to the empty flask. Reagent chemicals should never be added directly to a dry glass surface, as glass is highly absorbent. When using a volumetric flask, a solution should be prepared stepwise. The desired reagent chemical (either solid or liquid) to be diluted is added to the flask, and then diluent is added to fill the flask about two-thirds (taking care to rinse down any reagent on the ground glass lip). It helps to swirl the solution before diluent is added to the neck of the flask to obtain most of the mixing (or dissolving in the case of a solid). Finally, diluent is added so that the bottom of the meniscus is even with the middle of the calibration mark (at eye level). If there are any droplets of water on the neck of the flask above the meniscus, take a piece of tissue and blot these out. Also, dry the ground-glass stopper joint.

The solution is finally thoroughly mixed as follows. Keeping the stopper on securely by using the thumb or palm of the hand, invert the flask and swirl or shake it vigorously for 5 to 10 s. Turn right side up and allow the solution to drain from the neck of the flask. Repeat at least 10 times.

Note. When preparing the solution of an expensive chemical, should the volume of liquid go over the calibration mark, it is still possible to save the solution as follows. Paste against the neck of the flask a thin strip of paper and mark on it with a sharp pencil the position of the meniscus, avoiding parallax error. After removing the thoroughly mixed solution from the flask, fill the flask with water to the calibration mark. Then by means of a buret or small volume graduated pipet, add water to the flask until the meniscus is raised to the mark on the strip of paper. Note and record the volume so added and use it to mathematically correct the concentration calculation. For an inexpensive chemical, start over. If the volume goes over the mark, you cannot accurately calculate concentration without determining how far over the mark you went. Be very careful and patient when filling volumetric flasks, especially when the components in the flask are irreplaceable or expensive.

Pipets

Pipets are designed to deliver a specified volume at a given temperature, and they are marked "TD." Again, the volume can be considered to be constant with small changes in temperature. Pipets are calibrated to account for the drainage film remaining on the glass walls. This drainage film will vary somewhat with the time taken to deliver, and usually the solution is allowed to drain under the force of gravity and the pipet is removed shortly after the solution is delivered. A uniform drainage time should be adopted.

The volumetric pipet is used for accurate measurements since it is designed to deliver only one volume and is calibrated at that volume. Accuracy to four significant figures is generally achieved, although with proper calibration, five figures may be obtained if necessary. See the table on the back cover for tolerances of class A transfer pipets. Measuring pipets are straight-bore pipets that are marked at different volume intervals. These are not as accurate because nonuniformity of the internal diameter of the device will have a relatively larger effect on total volume than is the case for pipets with a bulb shape. Also, the drainage film will vary with the volume delivered. At best, accuracy to three significant figures can be expected from these pipets, unless you make the effort to calibrate the pipet for a given volume delivered.

Volumetric Glassware: Also Indispensable (figure 2)

Most volumetric pipets are calibrated to deliver with a certain small volume remaining in the tip. This should not be shaken or blown out. In delivering, the pipet is held vertically and the tip is touched on the side of the vessel to allow smooth delivery without splashing and so that the proper volume will be left in the tip. The forces of attraction of the liquid on the wall of the vessel will draw out a part of this.

Some pipets are blowout types (including measuring pipets calibrated to the entire tip volume). The final volume of solution must be blown out from the tip to deliver the calibrated amount. These pipets are easy to identify, as they will always have one or two ground bands or rings around the top. (These are not to be confused with a colored ring that is used only as a color coding for the volume of the pipet.) The solution is not blown out until it has been completely drained by gravity. Blowing to increase the rate of delivery will change the volume of the drainage film.

Volumetric pipets are available in sizes of 100 to 0.5 mL or less. Measuring and serological pipets range from a total capacity of 25 to 0.1 mL. Measuring pipets can be used for accurate measurements, especially for small volumes, if they are calibrated at the particular volume wanted. The larger measuring pipets usually deliver too quickly to allow drainage as fast as the delivery, and they have too large a bore for accurate reading.

In using a pipet, one should always wipe the outside of the tip dry after filling. If a solvent other than water is used, or if the solution is viscous, pipets must be recalibrated for the new solvent or solution to account for difference in drainage rate.

Pipets are filled by suction, using a rubber pipet bulb, a pipet pump, or other such pipetting device. Before using a pipet, practice filling and dispensing with water. No solution should be pipetted by mouth.

Syringe Pipets

These can be used for both macro and micro volume measurements. The calibration marks on the syringes may not be very accurate, but the reproducibility can be excellent if automatic delivery is used, such as a spring-loaded device that draws the plunger up to the same preset level each time. The volume delivered in this manner is free from drainage errors because the solution is forced out by the plunger. The volume delivered can be accurately calibrated. Microliter syringe pipets are used for introduction of samples into gas chromatographs. A typical syringe is illustrated in Figure 2.7. They are fitted with a needle tip, and the tolerances are as good as those found for other micropipets. In addition, any desired volume throughout the range of the syringe can be delivered. Syringes are available with a total volume as small as 0.5 μL, in which a wire plunger travels within the needle and so the entire syringe volume is within the needle.

Volumetric Glassware: Also Indispensable (figure 3)

The above syringe pipets are useful for accurate delivery of viscous solutions or volatile solvents; with these materials the drainage film would be a problem in conventional pipets. Syringe pipets are well suited to rapid delivery and also for thorough mixing of the delivered solution with another solution as a result of the rapid delivery.

A second type of syringe pipet is that shown in Figure 2.8. This type is convenient for rapid, one-hand dispensing of fixed (or variable) volumes in routine procedures and is widely used in the analytical chemistry laboratory. It contains a disposable nonwetting plastic tip (e.g., polypropylene) to reduce both film error and contamination. A thumb button operates a spring-loaded plunger, which stops at an intake or a discharge stop; the latter stop is beyond the former to ensure complete delivery. The sample never contacts the plunger and is contained entirely in the plastic tip. These pipets are available in volumes of 0.1 to 5000 μL and are reproducible to 1 to 2% or better, depending on the volume. Variable volume pipettes of this type typically deliver volumes across a defined range, such as 0.1 to 10 μL, 20 to 200 μL, 100 to 1000 μL, and 500 to 5000 μL. It is important to be consistent in your use of different pipettes for related procedures (e.g., always use a single draw from the 20 to 200 μL pipette to dispense 100 μL) to maintain consistency.

Sometimes, the actual volume delivered by these and other micropipets does not need to be known because they are used in relative measurements. For example, the same pipet may be used to deliver a sample and an equal volume of a standard solution for calibrating the instrument used for the measurement. Precision in delivery is usually more important than the absolute volume delivered. The European standard for pipet calibration in Europe is the German DIN 126650 (or a similar international standard ISO 8655). Calibrations are based on gravimetric testing (weighing of water). The DIN standard does not give separate limits for accuracy and precision, but rather uses a combined error limit equal to percent accuracy plus 2 times the standard deviation, that is, it gives a range in which we are 95% confident the delivery will fall.

Burets

A buret is used for the accurate delivery of a variable amount of solution. Its principal use is in titrations, where a standard solution is added to the sample solution until the endpoint (the detection of the completion of the reaction) is reached. The conventional buret for macrotitrations is marked in 0.1-mL increments from 0 to 50 mL; one is illustrated in Figure 2.9. The volume delivered can be read to the nearest 0.01 mL by interpolation, actually good to about ±0.02 or ±0.03 mL. Burets are also obtainable in 10-, 25-, and 100-mL capacities, and microburets are available in capacities of down to 2 mL, where the volume is marked in 0.01-mL increments and can be estimated to the nearest 0.001 mL. Ultramicroburets of 0.1-mL capacity graduated in 0.001-mL (1-μL) intervals are used for microliter titrations.

Volumetric Glassware: Also Indispensable (figure 4)

Drainage film is a factor with conventional burets, as with pipets, and this can be a variable if the delivery rate is not constant. The usual practice is to deliver at a fairly slow rate, about 15 to 20 mL per minute, and then to wait several seconds after delivery to allow the drainage to "catch up." In actual practice, the rate of delivery is only a few drops per minute near the end point, and there will be no time lag between the flow rate and the drainage rate. As the endpoint is approached, fractions of a drop are delivered by just opening, or "cracking," the stopcock and then touching the tip of the buret to the wall of the titration vessel. The fraction of the drop is then washed down into the solution with distilled water.

Care and Use of Volumetric Glassware

We have mentioned a few precautions in the use of volumetric flasks, pipets, and burets. Your laboratory instructor will supply you with detailed instructions in the use of each of these tools. A discussion of some general precautions and good laboratory technique follows.

Cleanliness of glassware is of the utmost importance. If films of dirt or grease are present, liquids will not drain uniformly and will leave water breaks or droplets on the walls. Under such conditions the calibration will be in error. Initial cleaning should be by repeated rinses with laboratory detergent and then water. Then try cleaning with dilute nitric acid and rinse with more water. Use of a buret or test tube brush aids the cleaning of burets and necks of volumetric flasks, but be careful of scratching the interior walls. Pipets should be rotated to coat the entire surface with detergent. There are commercial cleaning solutions available that are very effective.

Pipets and burets should be rinsed at least twice with the solution with which they are to be filled. If they are wet, they should be rinsed first with water, if they have not been already, and then a minimum of three times with the solution to be used; about one-fifth the volume of the pipet or buret is adequate for each rinsing. A volumetric flask, if it is wet from a previously contained solution, is rinsed with three portions of water only since later it will be filled to the mark with water. It need not be dry.

Note that analytical glassware should not be subjected to the common practice employed in organic chemistry laboratories of drying either in an oven (this can affect the volume of calibrated glassware) or by drying with a towel or by rinsing with a volatile organic solvent such as acetone (which can cause contamination). The glassware usually does not have to be dried. The preferred procedure is to rinse it with the solution that will fill it.

Care in reading the volume will avoid parallax error, that is, error due to incorrect alignment of the observer's eye, the meniscus, and the scale. Correct position is with your eye at the same level as the meniscus. If the eye level is above the meniscus, the volume read will be smaller than that taken; the opposite will be true if the eye level is too low.

After glassware is used, it can usually be cleaned sufficiently by immediate rinsing with water. If the glassware has been allowed to dry, it should be cleaned with detergent. Volumetric flasks should be stored with the stopper inserted, and preferably filled with distilled water. Burets should be filled with distilled water and stoppered with a rubber stopper when not in use.